← Biochemistry Biochemistry · Chapter 1

Laboratory foundations

Four everyday skills behind every biochemistry experiment: making dilutions, measuring concentration with light, controlling pH with buffers and judging replicates with simple statistics. Get these right and later results become trustworthy.

1. Read each section and learn the essentials by heart: they are the formulas you will use in every practical.
2. Open the 3D lab of the section, move one slider at a time and check that the plot changes the way the text predicts.
3. Test yourself with the 30-question quiz and redo the calculations on paper before checking the answer.
Overview

Chapter map

The chapter in one picture. Click a box to jump to its part.

LaboratoryfoundationsDilutionsC = C0 / f^nC1V1 = C2V2UV-VisA = ε · l · cread inside standardsBufferspH = pKa + log(B/A)useful at pKa ± 1StatisticsSE = SD / √nCI = mean ± t·SE
  1. DilutionsC = C0 / f^n · C1V1 = C2V2
  2. UV-VisA = ε · l · c · read inside standards
  3. BufferspH = pKa + log(B/A) · useful at pKa ± 1
  4. StatisticsSE = SD / √n · CI = mean ± t·SE
Part 1

Pipetting and serial dilutions

The essentials

  • Dilution factor f = final volume / aliquot volume (100 µL into 900 µL gives f = 10).
  • After n identical steps: Cₙ = C₀ / fⁿ; the total factor is the product of all steps.
  • For a single dilution, amount is conserved: C₁V₁ = C₂V₂.
  • Choose a calibrated pipette whose working range covers the volume you transfer.
  • Mix well at every step: an unmixed tube carries its error into every later tube.

A dilution lowers concentration by adding solvent while the amount of solute stays the same. The dilution factor f is the final volume divided by the volume of sample transferred. Transferring 100 µL of stock into 900 µL of buffer gives 1000 µL in total, so f = 10 (often written 1:10).

In a serial dilution the same step is repeated: tube 1 feeds tube 2, tube 2 feeds tube 3, and so on. Factors multiply, so after n steps the concentration is C0/fn. Three tenfold steps give a total factor of 1000. Serial dilutions are used to build calibration standards and to bring very concentrated samples into a measurable range without pipetting tiny volumes.

Stock 100 mM, three 1:10 steps: 100 / 103 = 0.1 mM. To make 10 mL of 2 mM from 100 mM: V1 = 2 × 10 / 100 = 0.2 mL stock + 9.8 mL buffer.

In the 3D lab the three sliders set the stock concentration, the dilution factor between adjacent tubes and the number of steps. Watch the liquid colour fade from tube to tube and the plot fall by the same factor at each step. Raising the factor makes the curve drop faster; adding steps adds tubes and plot points. The model assumes perfect mixing and perfect volumes.

In plain words: every step divides by the same number, so you multiply the numbers to know the total.

Interactive 3D model: move the sliders and watch what changes.

Part 2

Spectrophotometry and the Beer-Lambert law

The essentials

  • Beer-Lambert: A = ε · l · c (ε molar absorption coefficient, l path length, c concentration).
  • Absorbance and transmittance: A = −log₁₀T, so A = 1 means 10% of the light passes.
  • A blank (buffer only) sets the zero and removes background absorbance.
  • Read unknowns only inside the range of the standards; dilute if the signal is too high.
  • Original concentration = (A / slope) × dilution factor.

A spectrophotometer sends light of one wavelength through a cuvette and measures how much comes out. Transmittance T is the fraction transmitted; absorbance A = −log10T. A = 1 means 10% of the light passes, A = 2 only 1%.

The Beer-Lambert law says A = ε · l · c: absorbance is proportional to concentration c, to the path length l (usually 1 cm) and to the molar absorption coefficient ε of the molecule at that wavelength. Proteins absorb at 280 nm (tryptophan and tyrosine), nucleic acids at 260 nm, NADH at 340 nm (ε = 6220 M−1cm−1).

In practice you measure a calibration curve: standards of known concentration give a straight line whose slope is ε·l. An unknown is converted with c = A / slope. The line is only valid inside the range of the standards; at high absorbance the detector sees very little light and linearity is lost, so you dilute the sample and multiply back by the dilution factor.

Slope 0.4 AU/mM, diluted sample reads 0.5 AU, dilution 3-fold: 0.5 / 0.4 = 1.25 mM in the cuvette, × 3 = 3.75 mM in the original sample.

In the 3D lab the sliders change the calibration slope, the absorbance of the unknown and the dilution factor. A higher absorbance darkens the liquid and dims the light reaching the detector. The dilution factor changes only the back-calculated original concentration, not the measured absorbance. A warning appears when the diluted sample falls outside the 0-4 mM standards.

Interactive 3D model: move the sliders and watch what changes.

Part 3

pH, buffers and Henderson-Hasselbalch

The essentials

  • pH = −log₁₀[H⁺]; one pH unit is a tenfold change in [H⁺].
  • Henderson-Hasselbalch: pH = pKa + log₁₀([base]/[acid]).
  • Equal acid and base gives pH = pKa; a 10:1 ratio gives pKa + 1, a 1:10 ratio gives pKa − 1.
  • A buffer works best within about pKa ± 1.
  • Capacity depends on the total buffer concentration, not only on the ratio.

A buffer is a mixture of a weak acid (HA) and its conjugate base (A−). Added H+ is taken up by A−; added OH− is neutralised by HA. The pH therefore changes much less than in pure water. Enzymes and cells need this, because charge, structure and activity of proteins depend on pH.

The Henderson-Hasselbalch equation links pH to the ratio of the two forms: pH = pKa + log10([A−]/[HA]). When the two are equal the log term is zero and pH = pKa. Each tenfold change in the ratio moves the pH by one unit. Outside roughly pKa ± 1 one form dominates and the buffer resists poorly, so you choose a buffer whose pKa is close to the pH you want: Tris (pKa about 8.1) for slightly basic work, phosphate (pKa2 about 7.2) near neutral, acetate (pKa about 4.8) for acidic work.

Acetate, pKa 4.76, base/acid = 10: pH = 4.76 + 1 = 5.76. Base/acid = 0.1: pH = 3.76.

Buffer capacity is how much acid or base the buffer can absorb before the pH shifts noticeably. At the same ratio, a 100 mM buffer resists ten times more than a 10 mM one.

In the 3D lab one slider sets the pKa (shifting the whole curve up or down) and the other the base/acid ratio (changing the mix of particles and moving the point along the logarithmic curve). Check that the point sits exactly at pKa when the ratio is 1.

Interactive 3D model: move the sliders and watch what changes.

Part 4

Replicates, mean, SD and confidence intervals

The essentials

  • Mean = sum of values / n; the SD measures the spread of single values (divide by n − 1).
  • Standard error of the mean: SE = SD / √n; quadrupling n halves SE.
  • 95% confidence interval = mean ± t · SE, with t from n − 1 degrees of freedom (2.776 for n = 5).
  • Biological replicates are independent samples; technical replicates are repeated reads of one sample.
  • Coefficient of variation CV = SD / mean × 100%.

One measurement says little: you need replicates. Biological replicates are independent samples (different cultures, animals, preparations) and capture real variability. Technical replicates are repeated measurements of the same sample and only check the method; they must not be counted as independent n.

The mean summarises the centre. The standard deviation (SD) summarises how far single values scatter; it is computed with n − 1 in the denominator. The SD does not shrink with more replicates. What shrinks is the standard error of the mean, SE = SD/√n: our estimate of the mean becomes more precise.

A 95% confidence interval is mean ± t · SE. The t value comes from the t distribution with n − 1 degrees of freedom: 2.776 for n = 5, 2.262 for n = 10, 2.093 for n = 20, 2.023 for n = 40 (it approaches 1.96 for large n). Small samples pay twice: a larger SE and a larger t.

Mean 10, SD 2, n = 5: SE = 2 / √5 = 0.894; half-width = 2.776 × 0.894 = 2.48; CI = 7.52 to 12.48. With n = 20: SE = 0.447, half-width = 2.093 × 0.447 = 0.94.

In the 3D lab the sliders set the sample mean, the sample SD and the number of replicates (5, 10, 20 or 40). Moving the mean shifts the dots and the interval together; a larger SD spreads the dots and widens the interval; more replicates keep the spread but narrow the interval. The scale stays fixed so you can compare widths directly.

Interactive 3D model: move the sliders and watch what changes.